ECUT 400S electrosurgical generator system perform in mono & bipolar working modes but with different power range.
The 7 integrated different monopolar and bipolar modes, tissue responding and automatic return electrode monitoring to give intuitive indication of potential burn risk, which enables highly efficient and safe surgical cutting and coagulation through all surgical procedures.
Isolated power output
Quantitative display of contact area for "Split" type of neutral electrode
Contact area indicator lamp
Adjustable alarm volume
Monopolar |
|
|
|
|
Cutting mode |
Max. power output (watts) |
Load (ohms) |
Open circuit Vp-p |
Crest factor |
Pure |
350W |
500 |
1,800 / 1,200v |
1.5 |
Blend1 |
300W |
500 |
2,700 / 2,000v |
1.9 |
Blend2 |
200W |
500 |
2,400 / 1,700v |
2.3 |
Blend3 |
150W |
500 |
2,900 / 3,000v |
3.1 |
Coagulation mode |
||||
Point |
100W |
500 |
4,600 / 3,900v |
4.5 |
Soft |
100W |
500 |
3,600 / 3,300v |
4.5 |
Bipolar |
|
|
|
|
Coagulation mode |
|
|
|
|
Standard |
50W |
100 |
500v |
1.5 |
Compact Design
The ECUT 400S benefiting from an structural design and light weight make it mountable on support tray of ceiling pendant for endoscopic application.
Control and measurement
Front panel of ECUT 400S consists of two differently sloped areas, 130 ° sloped "Measurement area" and 160 ° sloped "Control area", to provide operator great physical comfort when they setup or adjust power setting and observe the power measurement and neutral electrode's contact situation from an ergonomic viewing angle and height.
Smoke evacuation (optional)
Heat-producing surgical methods (like electrosurgery or laser) mostly bring surgical smoke that contain hazardous (like acrylonitrile and hydrogen cyanide) which may blur surgical vision (especially for procedures into cavity) and cause harm to respiratory systems of patients, surgeons or operating theatre staff.
Auto-power
Resistance from nerve, muscle, adipose tissue and bone to electrical circuits vary. Impedance response technology in ECUT 400S electrosurgical system measures the tissue resistance then automatically adjust power output under pre-set range, to optimize real cutting or coagulation performance, which makes ECUT 400S system qualified for wide range of applications, from various open surgeries to minimal-access procedures, as well as under-water operation in Gynaecology and Urology.
Controls and Switches
Electrodes
Patient return electrode
Other accessories