Marcom 0.5T
Open MRI system
Marcom 0.5T is an open 0.5T permanent magnet MRI system equipped with powerful RF and gradient system, together with advanced imaging technology, making it a high-end system which is comparable to high-field MRI.
Features
Eddy zero technology
The shimming algorithm technology (ensures field strength with high uniformity and stability), ensure the MRI system operating stable with high quality and high performance.
Eddy Zero Technology:
-
Magnet design uses breakthrough anti-eddy current technology
-
The gradient coil using self-shielded active anti-eddy current technology
-
Implementation of precision eddy current compensation algorithm, completely eliminate the impact of eddy current.
The eddy current because of the high gradient and high slew rate come from ultra-fast sequence was resolved completely, which guarantee to get the best clinic image.
Open wide design
More patient comfort and more space for surgery. Affinity C-shape magnets, creating a maximum of openness, offer a maximum vision for patients, especially leave much space comfort for obese patients, minimizing claustrophobia of patients and providing more space for MR intervention surgery.
Advance gradient system
The gradient system helps to provide higher resolution pictures. The Marcom 0.5T (X, Y, Z) gradient intensity is 25mT/m, providing higher gradient intensity, faster imaging speed, shorter scanning time, thinner imaging slice and higher image resolution.
The new technology of 4D shimming
Use of advanced active shimming algorithm for real-time automatic shimming on each examination to ensure the magnetic field always maintain the highest uniformity. Magnetic field homogeneity and stability of the MRI images always are the most important guarantee of the high-resolution, high SNR and high contrast. They are the important indicators of the level of magnet design, the better the smaller the value. It directly determines the SNR of the image, like a car chassis, where stability is essential. The magnetic field uniformity is maintained at < 2.5 ppm in 40 cm the DSV Vrms, allowing the system to complete a wide range of scanning (40c m).
Advanced RF system
The Marcom 0.5T is equipped with fast 4 channels RF system and all phased array coils providing best SNR pictures.
Coils
Marcom 0.5T has a variety of different phased array coils and all of them provide the best SNR pictures.
Standard
- Head coil
- Neck coil
- Small body coil
- Large body coil
Optional
- Shoulder coil
- Sport joint coil
- Wrist Coil
- Breast Coil
- Flexible coil
- Flexible body coil
- Flat spine coil
- Knee coil
Outstanding
- Fully open Magnet
- Nd-Fe-B magnet
- 4D shimming
- Eddy Zero Technology
- self-regulating constant temperature
- Fully Digital 4 Channel Receiving Spectrometer
- Automatic coil tuning
- Accurate position assist
- Higher SNR
- Higher resolution
- Less Scanning Time
- Comprehensive scanning sequences
- Advanced imaging techniques
Specifications
- Permanent magnet with automatic constant temperature system
- Field strength 0.5T
- Dual-column shape magnet
- Homogeneity ≤2.5ppm
- Shimming method Active/Passive/Dynamic
- Optimal algorithm of active shimming
- Magnet vertical gap 41cm
- Accessibility (Horizontal opening angle) >274°
- Gradient field strength (Single Axis) 25mT/m
- Gradient slew rate (Single Axis) 75mT/m/ms
- FOV 20 to 410mm
- 2D thickness 1mm
- 3D thickness 0.1mm
- Air cooling gradient system
- Fully digital Spectrometer
- Digital Transmit and Receive RF system
- 4 channels RF system
- Phase array coil
- Dual-voice-patient communication system
- patient weight 200Kg
- Patient bed laser location system
- DICOM 3.0
Pulse sequences
- Spin-echo sequence
- GRE sequence
- IR sequence
Advance imaging technology
- Body Imaging
- MR Angiography
- Diffusion weighted imaging (DWI)
- Flow compensation
- Gating technology
- Pre-saturation technology (PS)
- Pre-saturation adjustment technology
- Part metal implant scan technology
- Multi-layer and multi-angle scanning technology
- Optimize bandwidth acquisition technology
- Parallel acquisition technology
- Section acquisition technology
- Oversampling technology
- Image fusion technology
- Artifact suppression technology
- Thin imaging technology
- Online image filtration
- Scan sequence queuing
- Movie playback technology
- Post processing package